4-Digit RTD Meter, Transmitter & Controller
Q9000M (0.1°)


    Electromagnetic Compatibility


NEWPORT PRODUCT INFO
MANUAL - PDF Version QxxxxM
MECHANICAL 1/8 DIN (1A) Case
MECHANICAL Connections Diagram
PRICE
REQUIRES ADOBE ACROBAT - HELP
QUANTA STANDARD FEATURES
100 ohm platinum RTD input
2,3 or 4-wire connection
±9,999-count display span
0.1°C or 0.1°F resolution
Sensor-break detection
120 dB CMR to 500 V
1 mV/deg linearized analog output
LED or LCD display
Bright, 0.56 in (14.2 mm) LED display
Display hold and test
120/240 Vac, 5 Vdc, 9-32 Vdc or 26-56 Vdc power
Screw-terminal barrier strip
1/8 DIN case
- -
OPTIONS
Analog output for user-selected span
Single-setpoint 10 A relay control
Dual-setpoint 10 A relay control
Proportional 4-20 mA control
Time-proportional 2 A solid-state relay control
Isolated parallel BCD output
NEMA-4 splash-proof lens cover

The QUANTA Q9000M is a 4-digit indicator/controller with a wide choice of output options. It provides 0.1 degree resolution for 100 ohm platinum RTD's with alpha of 0.00385. It is complete with sensor excitation and break detection. Readout can be in °C or °F, as selected by plug-in jumpers.

POWER OPTIONS
Six types of meter power are available: 120 Vac, 240 Vac, 24 Vac, 5 Vdc, isolated 9-32 Vdc, or isolated 26-56 Vdc.

ANALOG OUTPUT OPTIONS
A 1 mV/degree (±1 V full-scale) analog output is standard and is suitable for driving a strip-chart recorder. An additional analog output can be provided by an optional vertical plug-in board. Available output signals are 0-5 Vdc, 0-10 Vdc, 0-1 mA (source or sink), and 4-20 mA (source or sink). The top and bottom of each output range can be scaled to fit a user-selected display span. One advantage in using the Q9000M as a 4-20 mA transmitter is that the current signal tracks the display, and hence is linearized.

CONTROL OUTPUT OPTIONS
Additional outputs can be provided by a horizontal upper board. Available options include single-setpoint control with one 10 A relay, dual-setpoint control with two 10 A relays, 4-20 mA proportional control (source or sink), time- proportional 2 A solid-state relay control, and isolated, parallel BCD output. For additional information, please refer to the QUANTA, Options, and Mechanical sections.

SPECIFICATIONS

ANALOG INPUT

Temperature RangeRTD type: 100-ohm platinum with alpha of 0.00385
Calibration: To DIN 43760, October 1980
Excitation current:
0.42 mA
Sensor connection: 2, 3 or 4 wires
Max lead resistance: 20 ohm for 2, 3 or 4-wire connection
Zero adjustment: ±5°C (±9°F)
Overvoltage protection: 15 Vp

NOISE REJECTION
NMR: 115 dB, 50/60 Hz
CMR: 120 dB
CMV: 1500 Vp per HV test, 354 Vp per IEC spacing

ACCURACY AT 25°C
Zero tempco (bridge-network error): ±0.05 deg/deg
Span tempco, 10°C to 40°C: ±0.006% of reading/°C
Lead resistance effect
2-wire connection: +2.6°C (4.7°F)/ ohm
3- or 4-wire connection: compensated to 20 ohm max

ANALOG-TO-DIGITAL CONVERSION
Input configuration: Differential, bipolar
Technique: Dual-slope, average-value
Polarity: Automatic
Signal integration period: 100 ms
Read rate: 2.5/s

DISPLAY
Type: 7-segment, red LED
Height: 0.56 in (14.2 mm)
Symbols: ­8.8.8.8
Decimal-point positions: Three positions selectable by jumpers behind lens or at connector
0 to 9,999 counts: Normal operation
10,000 to 19,999 counts: Four least-significant digits flash correct reading
Beyond 19,999 counts: All four digits flash zeros
Sensor-break indication: All four digits flash zeros

POWER
AC voltages:
120, 240 or 24 V ac +10%/­15%
AC frequency: 49-440 Hz
DC voltages: 9-32 Vdc, isolated to 300 Vp; 26-56 Vdc, isolated to 300 Vp; 5 Vdc ±5%, non-isolated
Power consumption: 8 W max

ENVIRONMENTAL
Operating temperature:
0 to 60°C
Storage temperature: ­40 to +85°C
Relative humidity: 95% at 40°C (non-condensing)

MECHANICAL
Dimensions: Newport DIN1A (1/8 DIN) case (see Mechanical section for drawings)
Weight: 17 oz (480 g)
Case material: 94V-0 UL-rated polycarbonate



To Order (Specify Model Number, see variations below in partnumber builder table)    
 Part Number   Availability   Price    Description    Qty.
 Q9000-MDF1 in-stock    $355.00     Quanta meter,4 digit display, red LED display, 120 Vac power, 1 mV per count analog output, Pt100 RTD input with 0.1° resolution, -328.0 to 999.9°F range (no control output)  
 Q9000-MDC1 in-stock    $355.00     Quanta meter,4 digit display, red LED display, 120 Vac power, 1 mV per count analog output, Pt100 RTD input with 0.1° resolution, -200.0 to 830.0°C range (no control output)  
 Q9020-MDF1 1 Week    $590.00     Quanta meter,4 digit display, red LED display, 120 Vac power, 0 to 10 Vdc analog output, Pt100 RTD input with 0.1° resolution, -328.0 to 999.9°F range (no control output)  
 Q9005-MDC1 1 Week    $480.00     Quanta meter,4 digit display, red LED display, 120 Vac power, 1 mV per count analog output, single setpoint 10 A relay, Pt100 RTD input with 0.1° resolution, -200.0 to 830.0°C range  
 
All dollar amounts on this site are shown in US currency.

 Part Number Builder
(1) (2) (3) (4) (5) (6) (7)
Q -
Option Descriptions
(1) Display Resolution and Case Style
  Select
9 for ±9,999 counts (4 digits)
8 for ±9,999 with INFINITY FAMILY BEZEL AND ADAPTER
(2) Display Type & Meter Power - LCD
  Select
0 for LED, 120 Vac (49/440 Hz) (BQ20 or BQ90)
1 for LCD, 120 Vac (49/440 Hz) (BQ21)
2 for LED, 240 Vac (49/440 Hz) (BQ22 or BQ92)
3 for LCD, 240 Vac (49/440 Hz) (BQ23)
4 for LED, 9-32 Vdc (isolated) (BQ24 or BQ94)
5 for LCD, 9-32 Vdc (isolated) (BQ25)
6 for LED, 5 Vdc** (BQ26 or BQ96)
7 for LCD, 5 Vdc** (BQ27)
8 for LED, 24 Vac (BQ28 or BQ98)
9 for LCD, 24 Vac (BQ29)
A for LED, 26-56 Vdc (isolated) (BQ2A or BQ9A)
B for LCD, 26-56 Vdc (isolated) (BQ2B)
(3) Analog Output Adders
  Select
0 for ±1 or ±2 V recorder output, non-isolated. (standard)
1 for 0-5 Vdc, non-isolated (BA01)
2 for 0-10 Vdc, non-isolated (BA01)
3 for 0-1 mA, source or sink, non-isolated (BA01)
4 for 4-20 mA, source or sink, non-isolated (BA01)
5 for 4-20 mA, sink (high-compliance), non-isolated. (BA02)
6 for 4-20 mA, ISOLATED (BA03)
(4) Control Output Adders
  Select
0 for None
1 for Dual-setpoint 10 A relay. (BDT1)
2 for Proportional 4-20 mA control, source or sink (BDP1)
3 for Time-proportional solid-state 2 A relay plus proportional 4-20 mA control, source or sink. (BDP2)
4 for BCD (isolated, latched, buffered, byte addressable). Includes 50-pin mass-termination connector. (BDD2)
5 for Single-setpoint 10 A relay (BDS1)
(5) Signal Conditioning Input Adders
  Select
MDC1 for RTD (Pt 100) input -200.0 to +830.0°C (BSCM)
MDF1 for RTD (Pt 100) input -328.0 to +999.9°F (BSCM)
(6) Display Color
  Select
Nothing(leave field blank) for Red Display (standard)
,G for Green Display
(7) Lens
  Select
Nothing(leave field blank) for Standard Lens (with Newport logo)
,BL for Blank Lens
 NOTE: All combinations may not be valid, check spec sheet for valid part numbers.


   
Keywords: 4-Digit RTD Meter, Transmitter & Controller

   Back to top

Sitemap | Order Status | Careers | About Us | ISO 9001 | Worldwide Affiliates | Terms & Conditions
Copyright ©2012, NEWPORT Electronics, Inc. All rights reserved.